- Best energy efficiency on the market
- Excellent temporal and spatial temperature accuracy
- USB connection for recording data
- Temperature range up to +300 °C

Binder Drying and Heating Chambers with Mechanical Adjustment | ED 56
The strengths of a BINDER ED series drying chamber include routine drying and sterilization tasks up to +300 °C. Thanks to the natural convection, all thermal processes which use this drying chamber are highly efficient. The ED series ensures fast, uniform drying.
Binder Drying and Heating Chambers with Mechanical Adjustment | ED 56 General Information
- Temperature range:
• +5°C above ambient temperature to
• +300 °C
- Up to 30% lower energy consumption compared to conventional units on the market
- High temperature accuracy thanks to APT.line™ technology
- Natural convection
- Controller with LCD display
- Electromechanical control of the exhaust air flap
- 2 chrome-plated racks
- Stackable
- Class 2 integrated independent adjustable temperature safety device (DIN 12880) with visual alarm
- Ergonomic handle design
- USB port for recording data
Data
Designation |
ED056-230V |
ED056UL-120V |
Article Number | 9010-0333 | 9010-0334 |
Option model | Standard | Standard |
Designation |
ED056-230V |
ED056UL-120V |
Temperature range | +5 °C above ambient temperature to 300 °C | +5 °C above ambient temperature to 300 °C |
Temperature uniformity at 150°C | 2.3 ± K | 2.3 ± K |
Temperature fluctuation at 150°C | 0.4 ± K | 0.4 ± K |
Heating up time to 150°C | 50 min | 50 min |
Recovery time after door was opened for 30 s at 150°C | 19 min | 19 min |
Air Change Data
Designation |
ED056-230V |
ED056UL-120V |
Air change rate at 100°C | 8 x/h | 8 x/h |
Electrical Data
Designation |
ED056-230V |
ED056UL-120V |
Rated Voltage | 230 V | 120 V |
Power frequency | 50/60 Hz | 60 Hz |
Nominal power | 1,05 kW | 1,15 kW |
Unit fuse | 6,3 A | 12,5 A |
Phase (Nominal voltage) | 1~ | 1~ |
Measures
Designation |
ED056-230V |
ED056UL-120V |
Interior volume | 57 L | 57 L |
Net weight of the unit (empty) | 39 kg | 39 kg |
Load per rack | 30 kg | 30 kg |
Permitted load | 70 kg | 70 kg |
Wall clearance back | 100 mm | 100 mm |
Wall clearance sidewise | 565 mm | 565 mm |
Internal Dimensions
Designation |
ED056-230V |
ED056UL-120V |
Width net | 560 mm | 560 mm |
Height net | 625 mm | 625 mm |
Depth net | 565 mm | 565 mm |
Designation |
ED056-230V |
ED056UL-120V |
Interior width | 360 mm | 360 mm |
Interior height | 420 mm | 420 mm |
Interior depth | 380 mm | 380 mm |
Unit doors | 1 | 1 |
Environment Specific Data
Designation |
ED056-230V |
ED056UL-120V |
Energy consumption at 100°C | 105 Wh/h | 105 Wh/h |
Energy consumption at 150°C | 180 Wh/h | 180 Wh/h |
Fixtures
Designation |
ED056-230V |
ED056UL-120V |
Number of shelves (std./max.) | 2/4 | 2/4 |