Benefits
- Stable and uniform environment
- Fully complies with Pharma regulations (ICH)Equipment
- Longevity - corrosion resistant
- Ease of operation
- Environmental friendly refrigerant with GWP value 547
- Maintenance possible with the least possible disruption to the climate conditioning
- Perfectly meets requirements in Pharma industry
- 3 years warranty

Binder Walk-in Chambers ICH (Q1A) | WIC 2
The walk-in-chambers offer a wide range of features and benefits to perfectly meet requirements in stability testing. Perfect test conditions are maintained in a wide range of temperature from 10°C to 50°C and relative humidity of 20% to 90% RH.

Binder Walk-in Chambers ICH (Q1A) | WIC 2 General Information
- Relative humidity accuracy of ± 2.5%
- Temperature accuracy of ± 1.5°C
- Duct work for proper distribution of air
- Highly insulated room panels with U value of only 0,21 W/m2K
- LED Lighting with motion detector
- High quality room elements, made in Germany
- Interior completely stainless steel
- Exterior service accessibility
- Room Elements: Galvanized sheet steel, polyester coated, white, similar to RAL 9010, interior surfaces in stainless steel,4 inch / 100 mm PU foam U value 0,21W/m2K
- Floor: Floor overlay is chrome-nickel steel, pattern-rolled 5WL
- Surface load: 50.000 N/m2
Binder Walk-in Chambers | ICH (Q1A) Technical Specifications
Data
| Designation |
WIC2-230V |
WIC2-208V |
WIC2C-208V |
| Article Number | 9800-0012 | 9800-0003 | 9800-0004 |
| Option model |
Standard |
Standard |
with conditioner on ceiling |
| Designation |
WIC2-230V |
WIC2-208V |
WIC2C-208V |
| Temperature range with humidity | 10...50 °C | 10...50 °C | 10...50 °C |
| Temperature uniformity at 25°C and 60% RH | 1.5 ± K | 1.5 ± K | 1.5 ± K |
| Temperature uniformity at 40°C and 75% RH | 1.5 ± K | 1.5 ± K | 1.5 ± K |
| Temperature fluctuation at 25°C and 60% RH | 0.5 ± K | 0.5 ± K | 0.5 ± K |
| Temperature fluctuation at 40°C and 75% RH | 0.5 ± K | 0.5 ± K | 0.5 ± K |
Electrical Data
| Designation |
WIC2-230V |
WIC2-208V |
WIC2C-208V |
| Rated Voltage | 230 V | 208...230 V | 208...230 V |
| Power frequency | 50 Hz | 60 Hz | 60 Hz |
| Nominal power | 2,7 kW | 2,7 kW | 2,7 kW |
| Unit fuse | 20,0 A | 20,0 A | 20,0 A |
| Phase (Nominal voltage) | 1~ | 1~ | 1~ |
| Designation |
WIC2-230V |
WIC2-208V |
WIC2C-208V |
| Interior volume | 18,523 L | 18,523 L | 18,523 L |
| Net weight of the unit (empty) | 981 kg | 981 kg | 981 kg |
| Designation |
WIC2-230V |
WIC2-208V |
WIC2C-208V |
| Width net | 2,600 mm | 2,600 mm | 2,600 mm |
| Height net | 2,470 mm | 2,470 mm | 2,470 mm |
| Depth net | 3,600 mm | 3,600 mm | 3,600 mm |
Internal Dimensions
| Designation |
WIC2-230V |
WIC2-208V |
WIC2C-208V |
| Interior width | 2,400 mm | 2,400 mm | 2,400 mm |
| Interior height | 2,270 mm | 2,270 mm | 2,270 mm |
| Interior depth | 3,400 mm | 3,400 mm | 3,400 mm |
| Unit doors | 1 | 1 | 1 |
| Designation |
WIC2-230V |
WIC2-208V |
WIC2C-208V |
| Energy consumption at 40°C and 75% RH | 1,140 Wh/h |
1,140 Wh/h |
1,140 Wh/h |
Binder Walk-in Chambers ICH (Q1A) | WIC 2 Images